Pma/510(k) #: exempt.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
Summary of event: as reported, an ngage nitinol stone extractor was damaged when removing the device from the package.A photo provided by the user appears to show the basket sheath having separated.The device did not make patient contact.A new device was used to complete the procedure.There was no impact to the patient as a result of this occurrence.Investigation evaluation: reviews of the complaint history, device history record, instructions for use (ifu), manufacturing instructions, and quality control procedures were conducted during the investigation.A visual inspection and functional test of the returned complaint device was also conducted.The complaint device was returned in in an open packaging tray.The collet and male luer lock adapter (mlla) were tight.The polyethylene terephthalate tubing (pett) was present.The basket sheath was damaged approximately 1 centimeter from the end of the male luer lock adapter (mlla).A review of the device history record found no non-conformances related to the reported failure mode.A review of complaint history records shows no other related complaints associated with the complaint device lot.Because there were no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that there is no evidence that nonconforming product exists in house or in the field.A review of relevant manufacturing and quality control documents was conducted.All extractors are verified to assure the basket opens and closes properly.Cook has concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.The instructions for use (ifu) provides the following information to the user related to the reported failure mode: important: excessive force could damage device.The returned device was found to be nonfunctional due to sheath damage.The basket sheath was severely kinked/buckled 1 centimeter from the distal end of the yellow support sheath.Based upon the available information and results of the investigation, the cause for the reported failure mode could not be conclusively determined.Excessive force may have been inadvertently applied to the device, but no information was provided regarding device handling.The appropriate personnel have been notified and cook will continue to monitor for similar events.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|