• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ACCU-PASS STR SHUTTLE 45 DEG RGHT CRVE; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ACCU-PASS STR SHUTTLE 45 DEG RGHT CRVE; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 7210424
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/20/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during an anterior cruciate ligament reconstruction, the accu-pass guide advances through the bearing located in the handle of the device and then it is passed to the surgeon, at the moment of penetrating the tissue, the guide advances in a short length and then stops advancing, the surgeon makes a strong movement inside the joint and the accu-pass at its tip tears the tissue, the procedure was completed with a s+n back up device.There was no delay and no further complications were reported.
 
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that during an anterior cruciate ligament reconstruction, the accupass guide advances through the bearing located in the handle of the device and then it is passed to the surgeon, at the moment of penetrating the tissue, the guide advances in a short length and then stops advancing, the surgeon makes a strong movement inside the joint and the accu-pass at its tip tears the tissue, the procedure was completed with a s+n back up device.There was no delay and no further complications were reported.The patient health is stable.
 
Additional Manufacturer Narrative
H10: internal complaint reference (b)(4).H2: additional information in b5.H3, h6: the reported device was received for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found no similar reported events.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A visual inspection revealed the device was returned outside of original packaging.A monofilament was preloaded into the device and a second monofilament was returned separately.No physical damage visible to the device.A functional evaluation revealed the device functions as intended.No relevant supporting clinical information has been provided to assist with a clinical investigation.Therefore, based on insufficient information, a thorough clinical assessment cannot be performed at this time.The patient's current condition was reported as stable.Should any additional clinical information be provided this complaint will be re-evaluated.There was no relationship found between the device and the reported event.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCU-PASS STR SHUTTLE 45 DEG RGHT CRVE
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14644932
Report Number1219602-2022-00847
Device Sequence Number5574573
Product Code LXH
UDI-Device Identifier03596010534453
UDI-Public03596010534453
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/14/2024
Device Model Number7210424
Device Catalogue Number7210424
Device Lot Number2065532
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/20/2022
Supplement Date Received by Manufacturer12/22/2022
Initial Report FDA Received Date06/09/2022
Supplement Report FDA Received Date12/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age20 YR
Patient SexMale
-
-