| Model Number |
61000 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The terumo bct technician evaluated the machine at the customer site.The iv pole was adjusted then verified to be operational.Investigation is in process.A follow up report will be provided.
|
| |
|
Event or Problem Description
|
|
The customer reported that the iv pole would not stay up.There was not a patient involved at the time of the iv pole falling, therefore patient information is not reasonably known.
|
| |
|
Event or Problem Description
|
|
The customer reported that the iv pole would not stay up.There was not a patient involved at the time of the iv pole falling, therefore patient information is not reasonably known.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being filed to provide additional information in h.6 and h.10.Investigation: the terumo bct technician evaluated the machine at the customer site.The iv pole was adjusted then verified to be operational.The device serial number history report indicates no further related issues have been reported for this device.One year of service history was reviewed for this device with no other problems identified related to the reported condition.Root cause: a root cause assessment was performed for this complaint.The root cause was determined to be related to component wear and tear.
|
| |
|