• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS Back to Search Results
Model Number TMICL13.2
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Codes Corneal Edema (1791); Intraocular Pressure Increased (1937); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/10/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Implant date, explant date:unk type of investigation - lens work order search: no similar complaint type events reported for units within the same lot.Claim#: (b)(4).
 
Event or Problem Description
The reporter indicated that a 13.2mm, tmicl13.2, -16.00/2.5/095 (sphere/cylinder/axis) implantable collamer lens was explanted due to sizing issues.Attempts to obtain additional information have not been successful.If additional information is received a supplemental medwatch report will be submitted.
 
Additional Manufacturer Narrative
B5-the reporter indicated that a 13.2mm, tmicl13.2, -16.00/2.5/093 (sphere/cylinder/axis) implantable collamer lens was implanted into the patient's left eye (os) on (b)(6) 2022.On (b)(6) 2022 the lens was explanted due to significant reduction of iridocorneal angles.For cause details the reporter stated "at the day 1 postoperative visit, the uncorrected visual acuity measured 20/320 os.Intraocular pressure was 45 mmhg os.Examination demonstrated severe corneal edema os with narrow angles ou." a replacement lens of shorter length was implanted on the same day as removal.If additional information is received a supplemental medwatch report will be submitted.H6- health effect- clinical code 4581: narrow angles.Claim # (b)(4).
 
Additional Manufacturer Narrative
Claim# (b)(4).
 
Additional Manufacturer Narrative
B5 - per email on (b)(6) 2023, problem was resolved.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene enriquez
1911 walker avenue
monrovia, CA 91016
MDR Report Key14645301
Report Number2023826-2022-01780
Device Sequence Number20063354
Product Code QCB
UDI-Device Identifier00841542109888
UDI-Public00841542109888
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 05/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2022
Device Model NumberTMICL13.2
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/11/2022
Supplement Date Received by Manufacturer07/05/2022
10/10/2022
06/21/2023
Initial Report FDA Received Date06/09/2022
Supplement Report FDA Received Date09/15/2022
10/10/2022
07/12/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/12/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age36 YR
Patient SexFemale
-
-