| Model Number |
1012536-80 |
| Medical Device Problem Codes |
Mechanical Jam (2983); Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure was to treat an external iliac artery.The 8.0x80mm absolute self-expanding stent system (sess) was attempted to be deployed; however, resistance was noted with the thumbwheel.The sess was removed without issue as the sess could not deploy; however, it is unknown if the device partially or completely failed to deploy.Another same size absolute was used to successfully complete the procedure.There was no adverse patient effects and no clinically significant delay.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record of the reported lot could not be conducted because the lot number was not provided.It is possible that the distal shaft was bent in the anatomy and/or interaction with other devices resulted in preventing the shaft lumens from moving freely resulting in the reported mechanical jam and the reported activation/deployment failure; however this cannot be confirmed.The investigation determined a conclusive cause for the reported difficulties cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|