| Catalog Number |
LX105 |
| Medical Device Problem Code |
Failure to Form Staple (2579)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch #: unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Photo analysis: this is an analysis of a photo submitted to ethicon endo surgery for review.During the visual analysis the following was observed: the photo shows a product lx105 code device apparently in good condition, without bodily fluids.Based on the set of photos review, the event describe is not confirmed, however no conclusion or root cause could be determined.Hands on device analysis may provide the additional evidence necessary to confirm the root cause of the reported event.Because the instrument was not returned our evaluation is limited.We value the opportunity to fully analyze the instrument upon its return.As part of ethicon endo surgery quality process all devices are manufactured, inspected, and released to approved specifications.The batch history records were reviewed and certed by external manufacturing that the manufacturing criteria was met prior to the release of the equipment.The certificate records are accessible through external manufacturing.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unknown procedure the clipper does not fully close the clip.The procedure was successfully completed and there were no patient consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 8/12/2022.Investigation summary the product was returned to ethicon endo surgery for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that the lx105 device was returned nonfunctional as the jaws were misaligned; therefore, the clips could not be loaded into the jaws.No conclusion could be reached as to what may have caused the found condition of jaws.Please note that as stated in the ifu: all surgical instruments are subject to a degree of wear and tear because of normal use.A regular and precise visual check of the instrument should be made before each use.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.
|
| |
|
Search Alerts/Recalls
|