Model Number 176620
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); Perforation of Vessels (2135)
Date of Event 05/23/2022
Type of Reportable Event Serious Injury
Event or Problem Description
According to the reporter, during a procedure, the surgeon described not feeling the clamping force of the clip when using the device.The surgeon had difficulty using the clip applier.The clip was incorrectly positioned clip twice, hitting the artery and caused a little bleeding (less than 500cc).The removal of the clips resulted in a breach of the cystic artery.A new clip applier was used to complete the procedure.There was no tissue damage and the patient is currently doing well.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Additional information: d9, g1 (manufacturer name, mfr contact first name, last name, street 1, mfr city, region, postal code, email, phone number), g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the clips did not close completely and were malformed.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: avoid excessive twisting of the jaws or tissue manipulation when firing the instrument deflecting the jaws and/or shaft during firing may result in an improperly formed clip and possible bleeding and/or leakage.Squeeze the handle firmly as far as it will go.As the handle is squeezed, the clip is held firmly by the jaw and closes around the tissue.Failure to squeeze the handle as far as it will go can result in clip malformation and possible bleeding or leakage.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO CLIP
Common Device Name
CLIP, IMPLANTABLE
Manufacturer (Section D)
CELESTICA DE MONTERREY S.A. DE C.V.
calle octava #102
parque industrial monterrey
apodaca nuevo leon
MX 
Manufacturer (Section G)
CELESTICA DE MONTERREY S.A. DE C.V.
calle octava #102
parque industrial monterrey
apodaca nuevo leon
MX  
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view,mn, CO 55112
7635265677
MDR Report Key14648557
Report Number3003139373-2022-00155
Device Sequence Number11672285
Product Code FZP
UDI-Device Identifier10884521057876
UDI-Public10884521057876
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K143644
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/16/2025
Device Model Number176620
Device Catalogue Number176620
Device Lot NumberT0F134X
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/25/2022
Supplement Date Received by Manufacturer08/17/2022
Initial Report FDA Received Date06/09/2022
Supplement Report FDA Received Date09/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-