Model Number CD3361-40QC
Device Problem Charging Problem (2892)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/23/2022
Event Type  malfunction  
Event Description
Related manufacturer report number 2017865-2022-12099.It was reported that during a remote follow up, fast charge time was noted on the device and a temporary high defibrillation impedance were noted on the right ventricular (rv) lead, suspected to have occurred during an unrelated epicardial lead implant procedure.Abbott technical support was contacted and the fast charge time and high defibrillation impedance were suspected to be due to the unrelated procedure.A capacitor maintenance was performed during a follow up in clinic and the charge time was within normal limits.No additional intervention was required.The patient was in stable condition and will continue to be monitored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY ASSURA ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key14648877
MDR Text Key294033910
Report Number2017865-2022-11896
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodePO
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/09/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2023
Device Model NumberCD3361-40QC
Device Lot NumberP000123230
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/23/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/19/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MEDTRONIC MODEL 4968 - 60CM; TENDRIL STS
-
-