MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPICC PROVENA SOLO 5F DL FT MAX BARRIER PLUS KIT; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPICC PROVENA SOLO 5F DL FT MAX BARRIER PLUS KIT; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported "introducer in kit had a lip at the tip that wouldn't allow to insert." no other information was provided.
 
Additional Manufacturer Narrative
Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The manufacturer has received the sample and is pending evaluation.Results are expected soon.A lot history review (lhr) of refz2047 showed no other similar product complaint(s) from this lot number.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, complaint and lot history review, applicable previous investigation(s), sample (if available), applicable manufacture records, and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a damaged microintroducer sheath is confirmed; however, the exact cause is unknown.One photograph of a microintroducer was returned for evaluation.An initial visual observation of the photograph showed the distal edge of the microintroducer sheath was damaged, torn, and buckled.No further detail of the damage could be seen in the returned photograph.What appears be blood residue was observed within the dilator of the microintroducer in the returned photograph.While the cause of the damage observed on the sample in the returned photograph could not be determined, possible contributing factors include insertion technique, advancement against resistance, insertion angle, and patient anatomy.H3 other text : evaluation findings are in section h11.
 
Event or Problem Description
It was reported "introducer in kit had a lip at the tip that wouldn't allow to insert." no other information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC PROVENA SOLO 5F DL FT MAX BARRIER PLUS KIT
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kayla olsen
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key14650457
Report Number3006260740-2022-02153
Device Sequence Number5693847
Product Code LJS
UDI-Device Identifier00801741155017
UDI-Public(01)00801741155017
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K072230
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberS1295108FD3
Device Lot NumberREFZ2047
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/18/2022
Supplement Date Received by Manufacturer07/08/2022
Initial Report FDA Received Date06/09/2022
Supplement Report FDA Received Date07/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-