C.R. BARD, INC. (BASD) -3006260740 FULL KIT 20G X 8 CM WITH GUARDIVA AND PROBE COVER; CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/17/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported "nurse shared that the powerglide inserted well in the vein and the wire deployed successfully however during the part of advancing the catheter resistance was met.After resistance was met the nurse removed the entire set with everything intact successfully and observed the catheter going through the needle and the wire being bundled up.".
 
Additional Manufacturer Narrative
The manufacturer has received the sample and is pending evaluation.Results are expected soon.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), photo analysis and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint that the catheter became damaged during the insertion procedure was confirmed but the cause is unknown.A single photograph of a damaged powerglide pro, with the deployment device, was returned for evaluation.The guidewire had been deployed but the catheter wings were in the non-deployed position.The catheter contained a split where the introducer needle was seen protruding through the catheter shaft.The catheter was bent upon itself with the distal end of the catheter no longer around the needle but with the guidewire still inlaid.A kink was noted in the guidewire which may be indicative of a difficult placement.The observable features on the submitted photographs did not contain enough information to identify a definitive root cause of the reported event.Possible contributing factors could include insertion technique, product dimensions, advancement of the catheter against resistance, or reinsertion of the needle.The ifu warns, ¿once the catheter has been advanced, do not re-insert the needle back into the catheter or pull the catheter back onto the needle.If the catheter needs to be repositioned, either do so without the aid of the needle, or remove both the catheter and the needle as a unit to prevent the needle from damaging or shearing the catheter.¿ h3 other text : evaluation findings are in section h11.
 
Event or Problem Description
It was reported "nurse shared that the powerglide inserted well in the vein and the wire deployed successfully however during the part of advancing the catheter resistance was met.After resistance was met the nurse removed the entire set with everything intact successfully and observed the catheter going through the needle and the wire being bundled up.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FULL KIT 20G X 8 CM WITH GUARDIVA AND PROBE COVER
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kayla olsen
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key14650575
Report Number3006260740-2022-02152
Device Sequence Number5693357
Product Code FOZ
UDI-Device Identifier00801741140570
UDI-Public(01)00801741140570
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K162377
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberF120088PT
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/18/2022
Supplement Date Received by Manufacturer07/05/2022
Initial Report FDA Received Date06/09/2022
Supplement Report FDA Received Date07/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age55 YR
Patient SexMale
Patient Weight87 KG
-
-