| Model Number |
CA |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
Tooth Fracture (2428)
|
| Date of Event |
04/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
After investigation, the scan sent by the doctor was not accurate due to a bad capture.A dimentional distortion of -0.1 mm was found on the original scan leading to a tight fit on the incisors.The device fitted the initial models without resistance.
|
| |
|
Event or Problem Description
|
|
The device was too tight on try in.The incisal edge of tooth #8 was fractured.
|
| |
|
Search Alerts/Recalls
|