• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER DENTAL IMP TM 3.7MM MTX,13MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER DENTAL IMP TM 3.7MM MTX,13MM; DENTAL IMPLANT Back to Search Results
Model Number TMMB13
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 04/22/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmerbiomet complaint number cmp-(b)(4).Weight unknown/not provided.Expiration date and unique identifier (udi) number unknown/not provided.Device manufacturer date unknown/not provided.
 
Event or Problem Description
It was reported that the implant could not be placed due to fracture of the patient's bone.The site was grafted for future placement.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
Zimmerbiomet complaint number (b)(4).The reported device was returned for investigation.Visual inspection of the returned product identified worn markings due to usage.No damage.Measurements were made to determine if the device matched print specification and it was confirmed to match the drawing specifications.Functional testing to recreate the reported event could not be performed due to the nature of the device and event.No pre-existing conditions were noted on the product experience report (per).The reported device was intended for tooth location #6 but could not be placed.Pictures or x-ray images were not provided.Complainant reported that the implant could not be placed due to fracture of the patient's bone.The reported complaint could not be verified.A definitive root cause for this complaint could not be determined.A device history record (dhr)review was performed and no related deviations or nonconformances were noted.Also, a complaint history search was performed using our complaint handling system and there were no related complaints for this product lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMP TM 3.7MM MTX,13MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key14652505
Report Number0002023141-2022-01450
Device Sequence Number11991789
Product Code DZE
UDI-Device Identifier00889024018884
UDI-Public(01)00889024018884(17)241013(10)1232413
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K132258
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/13/2024
Device Model NumberTMMB13
Device Catalogue NumberTMMB13
Device Lot Number1232413
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/27/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/27/2022
Supplement Date Received by Manufacturer11/16/2022
Initial Report FDA Received Date06/10/2022
Supplement Report FDA Received Date11/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/15/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
Patient SexFemale
-
-