WELCH ALLYN INC. PRO 6000 W/LARGE CRADLE; THERMOMETER, ELECTRONIC, CLINICAL Back to Search Results
Model Number 06000-300
Medical Device Problem Code Excessive Heating (4030)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer reported that the probe tip of their braun pro6000 overheated and displayed inaccurate temperatures.There were no allegations of injury.The braun/welch allyn thermoscan pro 6000 ear thermometer is indicated for the intermittent measurement of human body temperature for patients having ages ranging from normal weight (full term) newborn to geriatric adults in a professional use environment.The probe cover is used as a sanitary barrier between the infrared thermometer and the ear canal.To help ensure accurate temperature measurements, the sensor itself is warmed to a temperature close to that of the human body.When the braun/welch allyn thermoscan is placed in the ear, it continuously monitors the infrared energy until a temperature equilibrium has been reached and an accurate measurement can be taken.Temperature regulation is part of a normal homeostasis where the body attempts to maintain an optimal body temperature.Temperature varies person to person secondary to both individual physiological factors, and environmental influences.Temperature is routinely obtained during routine physical assessment and is considered one part of a set of vital signs (blood pressure, pulse and respirations).Temperature can be assessed numerically through use of a thermometry device or also through patient presentation (patient verbalizations of coldness or hotness, presence or absence of shivering or sweating, skin color, and skin temperature to touch).During patient physical assessment if a thermometer malfunction occurs a back-up thermometer would likely be available for immediate use.In addition, trained clinicians would likely be able to determine if a patient has an acute variation in temperature based on presentation/lack of symptoms.In the event a treatment was rendered/not rendered based on an inaccurate temperature the malfunction is not likely to result in harm.Should the malfunction of inaccurate temperature readings recur, it is unlikely that serious injury or death would occur for the reasons stated above.Therefore hillrom does not consider this complaint involving inaccurate thermometer readings as reportable.Hillrom/welch allyn¿s investigation of pro 6000 ¿hot tip¿ complaints confirmed that aggressive cleaning and decontamination practices can cause liquid ingress.Consequently, liquid ingress can adversely affect the temperature sensor causing the pro 6000 device to behave inconsistently and to overheat the speculum tip.It is believed that replication of the malfunction with the returned devices from the customer at the manufacturing site has not been possible as any fluid that may have ingressed had likely evaporated, therefore not showing the ¿hot tip¿ malfunction.Based on hillrom¿s ability to replicate the malfunction of a ¿hot tip¿ on new devices that can potentially go above the built in risk mitigations of a safety cut off that could potentially cause a more serious injury we have deemed this complaint to be reportable.
 
Event or Problem Description
The customer reported that the probe tip of their braun pro6000 overheated and gave inaccurate temperature readings.There were no allegations of injury.This event was captured under hillrom ref # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRO 6000 W/LARGE CRADLE
Common Device Name
THERMOMETER, ELECTRONIC, CLINICAL
Manufacturer (Section D)
WELCH ALLYN INC.
4341 state street
skaneateles falls NY 13153
Manufacturer Contact
keighley crosthwaite
1069 state route 46 east
batesville, IN 47006
8129310130
MDR Report Key14652535
Report Number1316463-2022-00087
Device Sequence Number4639595
Product Code FLN
UDI-Device Identifier00732094309027
UDI-Public732094309027
Combination Product (Y/N)N
Initial Reporter CountryGM
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Remedial Action Notification
Type of Report Initial
Report Date (Section B) 06/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number06000-300
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/25/2022
Initial Report FDA Received Date06/10/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse