| Model Number |
N/A |
| Medical Device Problem Codes |
Overheating of Device (1437); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Initial reporter name and address: hospital name: (b)(6).The results of the investigation are as follows: review of the most recent repair record determined the cart was giving an overheating error; the power inlet module was burnt and would not power on.The power inlet module was replaced and resolved the reported issue.The event is confirmed.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported this cart had an overheat error.The event timing after surgery.There was no patient involvement.During product evaluation, it was discovered that the power inlet module was burnt and would not power on.No adverse events were reported as a result of this malfunction.No further information is available.
|
| |
|