• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I 3I T3® WITH DCD® TAPERED IMPLANT 5/4 X 13MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I 3I T3® WITH DCD® TAPERED IMPLANT 5/4 X 13MM; DENTAL IMPLANT Back to Search Results
Model Number BNPT5413
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet complaint number: (b)(4).Patient weight not provided.Event date not provided.
 
Event or Problem Description
It was reported that the implant was removed due to pain at the site.The implant was wiggling.Tooth #9.
 
Additional Manufacturer Narrative
This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: type of investigation codes were updated: 3331.H6: investigation findings code was updated: 213.H6: investigation conclusion code was updated: 4315.H10: narrative/data was updated.The reported item was returned for investigation.Visual inspection of the returned product identified worn markings due to usage, however no damage identified.Functional testing to recreate the reported event could not be performed due to the nature of the device and event.Measurements were made to determine if the device matched print specifications and confirmed.The reported devices were located on tooth #9 and was used for approximately 1 month.Pictures or x-ray images were not provided.Dhr review was completed for the subject lot number (2021060508).It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Sterilization record (op0210) was reviewed and verified to have passed all sterilization activities with no issues or nonconformities identified.Complaint history review was performed for the reported lot number (2021060508) for similar event and no other complaint was identified.Sep post market trending was reviewed and there were no actionable events or corrective actions for the reported event or device.Based on the available information, device malfunction did not occur, and the reported event was non-verifiable.
 
Event or Problem Description
No additional event information was received at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3I T3® WITH DCD® TAPERED IMPLANT 5/4 X 13MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key14652987
Report Number0001038806-2022-00886
Device Sequence Number4868237
Product Code DZE
UDI-Device Identifier00844868029489
UDI-Public(01)00844868029489(17)260623(10)2021060508
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K122300
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date06/23/2026
Device Model NumberBNPT5413
Device Catalogue NumberBNPT5413
Device Lot Number2021060508
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/24/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2022
Supplement Date Received by Manufacturer11/10/2022
Initial Report FDA Received Date06/10/2022
Supplement Report FDA Received Date11/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/23/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age79 YR
Patient SexMale
-
-