• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THORATEC CORPORATION HEARTMATE II LVAS IMPLANT KIT (WITH SEALED GRAFTS); VENTRICULAR (ASSIST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC CORPORATION HEARTMATE II LVAS IMPLANT KIT (WITH SEALED GRAFTS); VENTRICULAR (ASSIST) BYPASS Back to Search Results
Model Number 106015
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Abscess (1690); Bacterial Infection (1735); Skin Infection (4544)
Event Date 04/06/2022
Event Type  Injury  
Manufacturer Narrative
No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is completed.
 
Event Description
It was reported that the patient was admitted to the hospital (b)(6) 2022 for a driveline infection with an episode of anterior chest wall abscess with extension to the driveline.It was noted that the patient had noticed that the site of debridement near the driveline exit site was not healing.They had an incision and drainage of the abscess on (b)(6) 2022.Intraoperative cultures at that time were negative, but superficial cultures were positive for 3 different species of coagulase-negative staphylococci (cons) and corynebacterium.There was a plan to treat with vancomycin, cefepime, and flagyl (metronidazole), and then transition to vancomycin and augmentin (amoxicillin/clavulanic acid).Given concern for the exposed driveline and a nonhealing wound, the patient went to operating room on (b)(6) 2022 for the replacement of their heartmate ii pump and exteriorization of the driveline through the left superior abdomen.Pending further culture data the patient would transition to potential home-going intravenous (iv) antibiotic therapy.
 
Manufacturer Narrative
The patient had recurrent driveline infection (mfr#2916596-2022-11708).The new pump remains in use supporting the patient.A correlation between the device and the report of infection could not be conclusively determined.Infection is listed in the instructions for use as a potential adverse event that may be associated with the use of heartmate ii left ventricular assist system.Infection has been previously investigated and will continue to be monitored through quality data reviews, which are conducted on production and post product signals to evaluate if products are conforming to product requirements.No further information was provided.The manufacturer is closing the file on this event.
 
Event Description
It was reported that the patient remained in hospitalization until (b)(6) 2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTMATE II LVAS IMPLANT KIT (WITH SEALED GRAFTS)
Type of Device
VENTRICULAR (ASSIST) BYPASS
Manufacturer (Section D)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer (Section G)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer Contact
bob fryc
6035 stoneridge drive
pleasanton, CA 94588
7818528204
MDR Report Key14654109
MDR Text Key294903573
Report Number2916596-2022-11322
Device Sequence Number1
Product Code DSQ
UDI-Device Identifier00813024011224
UDI-Public00813024011224
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P060040
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/10/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date01/31/2019
Device Model Number106015
Device Catalogue Number106015
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/27/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/03/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age78 YR
Patient SexMale
Patient Weight99 KG
-
-