| Model Number |
NSLX145C |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch #: v9669k.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformance were identified.
|
| |
|
Event or Problem Description
|
|
It was reported that during a benign hysterectomy the tip of the shaft broke off.All fragments were retrieved and a new device was used to complete the case with no patient consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) date sent: 7/5/2022 d4 batch #: v9669k investigation summary the product was returned to ethicon endo-surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the nslx145c device was returned with the distal guide weld broken.In addition, the jaws were not damaged as reported.The device was tested on the generator and passed all functional testing.The energy output delivered from the device was verified.All tones were heard during functional testing.The damage to the distal guide was not enough to prevent the jaws to open and close as expected.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances related to the reported complaint condition were identified.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reached on the cause of the reported event.
|
| |
|
Search Alerts/Recalls
|