Catalog Number RBYPOD6
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has been returned and the investigation results are pending.A follow up mdr will be submitted upon completion of the device investigation.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the splenic artery using a lantern delivery microcatheter, pod coils, a non-penumbra catheter, and a non-penumbra guide sheath.During the procedure, the physician advanced the lantern into the catheter and reported that it became stuck inside.Therefore, the lantern was removed.A new lantern was used to successfully implant a pod coil into the target location.While advancing the next pod coil, it unintentionally detached in the lantern hub.Therefore, the lantern containing the pod coil was removed.The procedure was completed using the same lantern, same catheter, and guide sheath to implant four ruby coils, three pod packing coils (pod pcs), two pod coils, and three non-penumbra coils.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Evaluation of the returned pod6 confirmed that the embolization coil was detached.Evaluation revealed that the pusher assembly was fractured near its proximal end and the pull wire was retracted from the fractured location on the pusher assembly.If the device is inadvertently advanced at an extreme angle during use, damage such as a fracture may occur.If the pusher assembly is fractured, the embolization coil will likely detach.Further evaluation of the device revealed kinks on the pusher assembly and offset coil winds on the embolization coil.This damage was likely incidental to the complaint.Penumbra devices are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POD
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key14654750
Report Number3005168196-2022-00276
Device Sequence Number4842053
Product Code HCG
UDI-Device Identifier00814548019611
UDI-Public814548019611
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K173614
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/11/2022
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberRBYPOD6
Device Lot NumberF96239
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/03/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/16/2022
Supplement Date Received by Manufacturer06/23/2022
Initial Report FDA Received Date06/10/2022
Supplement Report FDA Received Date07/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/12/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age77 YR
Patient SexFemale
Patient Weight56 KG
Date Report Sent to FDA01/01/2005
-
-