MAUDE Adverse Event Report: ABBOTT VASCULAR 0.014 HI-TORQUE (HT) TURNTRAC GUIDE WIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR 0.014 HI-TORQUE (HT) TURNTRAC GUIDE WIRE Back to Search Results
Model Number 1020013
Medical Device Problem Codes Break (1069); Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the procedure was to treat a left anterior descending artery.The turntrac guide wire was advanced to the lesion without issue.However, it was noted that the guide wire unraveled.It was decided to be removed, and although resistance during removal was met with an unspecified source, the guide wire was able to be removed.Another guide wire was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was returned for analysis.The reported unraveled wire was able to be confirmed.The reported difficult to remove was unable to be replicated in a testing environment as they were based on operational circumstances.A review of the lot history record and similar incident review of the reported lot could not be conducted because the lot number was not provided.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that the reported difficulties (unraveled wire and difficult to remove) were a result of manipulation of guide wire in the anatomy and/or advancing/removing against resistance that ultimately resulted in the reported/noted device damages.There is no indication of a product quality issue with respect to manufacture, design or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
0.014 HI-TORQUE (HT) TURNTRAC GUIDE WIRE
Common Device Name
GUIDE WIRE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key14654887
Report Number2024168-2022-06243
Device Sequence Number5550690
Product Code DQX
UDI-Device Identifier08717648222702
UDI-Public08717648222702
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K173795
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1020013
Device Catalogue Number1020013
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/25/2022
Supplement Date Received by Manufacturer07/28/2022
Initial Report FDA Received Date06/10/2022
Supplement Report FDA Received Date08/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-