| Catalog Number |
337.88 |
| Medical Device Problem Code |
Disconnection (1171)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
05/16/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A physician reported that the tip of an ophthalmic backflush device fell off in the patient's eye during a vitrectomy procedure.The eye is filled with silicone oil.The physician indicated that the tip might come out when the silicone oil would be removed from the eye.The current patient prognosis was good.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
No sample has been received at manufacturing for evaluation.A review of the device history record (dhr) traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The root cause cannot be identified conclusively because no picture or sample has been received.This complaint has been reviewed and future data will be monitored for evidence of adverse trending and further action will be taken, as appropriate.At a minimum, this will include completing reviews of complaint class report levels on a monthly basis.The manufacturer internal reference number is: (b)(4).
|
| |
|