SYNTHES GMBH CHUCK/KEYLESS/DRILL SPEED FOR TRS; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
| Catalog Number |
05.001.208 |
| Medical Device Problem Codes |
Unintended System Motion (1430); Complete Loss of Power (4015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/06/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Additional narrative: device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition of the device being frozen/not moving identified during service and evaluation was confirmed.The assignable root cause was determined to be traced to component failure from wear.Udi: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported by russia that during service and evaluation, it was determined that the keyless chuck device was frozen/would not move.It was further observed that the device had illegible etch, color band chipping/fading, corrosion/rusting/pitting, and cosmetic damage due to worn etch.It was determined that the moving parts of the device did not move smoothly and the device did not function.It was further determined that the device failed pretest for general condition, marking and labeling, checking for mechanical free movement, and checking general function in running mode.It was noted in the service order that the device did not work.It was reported there were no delays to the surgical procedure and the surgery was completed as intended.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed, and it was determined that the device also had worn sear and corroded bearing.The assignable root cause was determined to be due to component failure from normal wear.
|
| |
|
Search Alerts/Recalls
|
|
|