SMITH & NEPHEW, INC. JRNY II XR BASEPLATE SZ 3 LT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 74022203
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 06/25/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that, after a tka surgery had been performed in (b)(6) 2019 with a journey ii xr system, patient experienced loosening in their left knee.A revision surgery was performed to explant the jrny ii xr baseplate sz 3 lt along with the insert to resolve the reported event.
 
Additional Manufacturer Narrative
The device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that based on the information provided in the case records forms, the root cause of the reported event(s) could not be definitively concluded and the patient impact beyond the reported loosening, revision, and termination of study participation cannot be determined.No further medical assessment could be rendered at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A review of the instructions for use for this device revealed that loosening, bending, cracking, or fracture of implant components can occur as a result of trauma, strenuous activity, improper alignment,or duration of service.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors and/or some potential probable causes that could contribute to the reported event have been identified as abnormal motion over time, bone degeneration, fit/sizing issue, lack of ingrowth, lifetime of device, and/or traumatic injury.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Additional Manufacturer Narrative
The device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, per complaint details, a left tka revision was performed approximately 2.5 years post implantation due to ¿loosening¿.The adverse event complaint record form (crf) documents the serious adverse event as unrelated to the study device and/or procedure; however, the adverse event medical board was unable to exclude that the study device had contributed to the reported issue with a start date of 25-jun-2020 and stop date/recovered/resolved of (b)(6) 2020 with termination of participation in the study after revision/replacement of tibial component with ¿cs¿ (which has not been clarified as of the time of this medical investigation; therefore it is unknown if this refers to a cement spacer).The crf showed zone 14 tibial baseplate with 1mm radiolucency at the 3 month follow-up with increasing radiolucency and additional zones at the 6 month checkup.Of note in the crf, the implant disposition shows the tibial base and insert were ¿removed¿/explanted, while the femoral and patellar components were retained.Based on the information provided in the crfs, the root cause of the reported event(s) could not be definitively concluded and the patient impact beyond the reported loosening, revision, and termination of study participation cannot be determined.No further medical assessment could be rendered at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for knee systems revealed that loosening, bending, cracking, or fracture of implant components can occur as a result of trauma, strenuous activity, improper alignment,or duration of service.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors and/or some potential probable causes that could contribute to the reported event have been identified as abnormal motion over time, bone degeneration, fit/sizing issue, lack of ingrowth, lifetime of device, and/or traumatic injury.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II XR BASEPLATE SZ 3 LT
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14655074
Report Number1020279-2022-02868
Device Sequence Number14840497
Product Code JWH
UDI-Device Identifier00885556615331
UDI-Public00885556615331
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K152726
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number74022203
Device Catalogue Number74022203
Device Lot Number18ESL0008
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2022
Supplement Date Received by Manufacturer07/08/2022
07/29/2022
Initial Report FDA Received Date06/10/2022
Supplement Report FDA Received Date07/11/2022
08/03/2022
Date Device Manufactured07/24/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
JRNY II XR INS XLPE MED 3-4 LT 8MM, LOT#:17LM18849.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-