• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES REAL INTELLIGENCE ROBOTIC DRILL GUARD; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES REAL INTELLIGENCE ROBOTIC DRILL GUARD; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number ROB10016
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/23/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, during cori assisted tka surgery, the real intelligence robotic drill guard broke, external to the patient.The piece fell in the drapes.Procedure was resumed, without any delay, with a back-up device.No patient complications were reported.
 
Additional Manufacturer Narrative
H3, h6: the real intelligence robotic drill guard, part number rob10016, used for treatment was not returned for evaluation.A relationship between the reported event and the device could not be established.A complaint history review for similar reported/confirmed complaints has identified prior events.While all products meet required manufacturing specifications prior to release a serial number, lot number, or part revision is required to link the device to a dhr or nc investigation.Refer to the real intelligence cori for knee arthroplasty user manual, section setting up a cori knee procedure: setting up the instruments, preparing the suction tubing for proper set up and handling.The failure mode and associated risk have been anticipated within the risk file.As a result of the risk assessment the quality team has been notified for further investigation.Although the reported problem was not confirmed through a visual or functional evaluation, factors contributing to the reported symptom may have been associated with an incomplete weld to secure the suction tube to the guard body.Based on the investigation, the need for a corrective action is not recommended or required at this time.Should any additional information be received the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
 
Additional Manufacturer Narrative
The manufacturer has identified that this event should be re-evaluated for mdr reporting.The detached component is of a size/shape that can be easily retrieved in its entirety and this malfunction has been resolved by using an available back-up device.Although this event may result in a short delay while the procedure is resumed and completed as intended, this malfunction has not caused or contributed, in the past, to a death or serious injury, nor has it been required a medical intervention to preclude permanent impairment of a body function or permanent damage to a body structure.This malfunction is unlikely to cause any death or serious injury if it were to recur, either.Therefore, this event is considered non-reportable pursuant to 21 c.F.R.§803.If further details are provided confirming the occurrence of a reportable event, our files will be updated accordingly, and a further report submitted outlining both the event details and our investigations performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE ROBOTIC DRILL GUARD
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14655099
Report Number3010266064-2022-00479
Device Sequence Number4868336
Product Code OLO
UDI-Device Identifier00885556757352
UDI-Public00885556757352
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K193120
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberROB10016
Device Catalogue NumberROB10016
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2022
Supplement Date Received by Manufacturer08/01/2022
08/01/2022
Initial Report FDA Received Date06/10/2022
Supplement Report FDA Received Date08/06/2022
12/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CORI ROBOTICS USA, SN: (B)(6).
Patient SexUnknown
-
-