ASCENSION ORTHOPEDICS, INC. RSS GLENOSPHERE INSERTER; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number HDL096006501
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 05/23/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, during reverse shoulder surgery, a glenosphere inserter broke inside the patient.The broken piece was buried in the implant and could not be retrieved.A finishing impactor was used instead to fully seat the implant.No delay was caused as a consequence of this problem.No further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
Section h3, h6: the product has not been received at the aus site for evaluation and photographs were not provided, so the complaint could not be confirmed.The following investigative actions were performed.Refer to the respective tasks for further detail.Complaint history review: the complaint history review identified no similar reported events for this device or for the same product lot.No adverse trend was identified.Future complaints for this failure mode will continue to be monitored and investigated as required.Risk management review (device): identified no issues which could have caused or contributed to the reported event.The failure mode was previously identified by the risk management file and the anticipated risk level is acceptable.Dhr/batch record review: identified no issues or manufacturing abnormalities which could have caused or contributed to the reported event.This review determined that the device met manufacturing specifications upon release for distribution.Capa/nc/pra/hhe/field action review: identified no previous events or issues which could have caused or contributed to the reported event.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.Device labeling/ifu review: identified no issues which could have caused or contributed to the reported event.This review determined that the device met labeling requirements upon release for distribution.Product prints/specifications/procedure review: identified no issues which could have caused or contributed to the reported event.This review determined that the device met product specifications upon release for distribution.As the device has not been received for evaluation, a definitive root cause could not be determined.The complaint alleges no injury to the patient nor a delay during surgery.It could not be determined whether the device contributed to the reported event.The rss glenosphere inserter is a reusable instrument expected to be used across multiple surgeries.The threaded tip of the instrument is screwed into the glenosphere in order to handle the implant during insertion.The tip of the instrument could break off if excessive bending force were applied to the inserter while assembled to the glenosphere.As the instrument is reusable and used during multiple surgeries, another possible cause is wear and tear to the tip during normal use.Based on this investigation, the need for corrective action is not indicated as no non-conformances or product deficiencies were identified and the risk level is acceptable.If additional information is later received, the complaint may be reopened.No further investigation is warranted for this complaint, though future complaints will be monitored and investigated as required.This investigation can be closed.
 
Additional Manufacturer Narrative
H3, h6: this complaint was re-opened as the product has been received at the aus site for evaluation.The complaint was confirmed.The following investigative actions were performed.Refer to the respective tasks for further detail.- complaint history review: the complaint history review identified no similar reported events for this device or for the same product lot.No adverse trend was identified.Future complaints for this failure mode will continue to be monitored and investigated as required.- risk management review (device): identified no issues which could have caused or contributed to the reported event.The failure mode was previously identified by the risk management file and the anticipated risk level is acceptable.- dhr/batch record review: identified no issues or manufacturing abnormalities which could have caused or contributed to the reported event.This review determined that the device met manufacturing specifications upon release for distribution.- capa/nc/pra/hhe/field action review: identified no previous events or issues which could have caused or contributed to the reported event.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.- device labeling/ifu review: identified no issues which could have caused or contributed to the reported event.This review determined that the device met labeling requirements upon release for distribution.- product prints/specifications/procedure review: identified no issues which could have caused or contributed to the reported event.This review determined that the device met product specifications upon release for distribution.- visual evaluation: visual examination of the part found that the threaded tip on the reusable instrument was broken off, the reported event was confirmed.Visual evaluation of the rss glenoid inserter found that the threaded tip on the reusable instrument was broken off, the reported event was confirmed.The rss glenosphere inserter is a reusable instrument expected to be used across multiple surgeries.The threaded tip of the instrument is screwed into the glenosphere in order to handle the implant during insertion.The tip of the instrument could break off if excessive bending force were applied to the inserter while assembled to the glenosphere.As the instrument is reusable and used during multiple surgeries, another possible cause is wear and tear to the tip during normal use.Based on this investigation, the need for corrective action is not indicated as no non-conformances or product deficiencies were identified and the risk level is acceptable.If additional information is later received, the complaint may be reopened.No further investigation is warranted for this complaint, though future complaints will be monitored and investigated as required.This investigation can be closed.H3, h6 (type of investigation, investigation findings, investigation conclusions).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSS GLENOSPHERE INSERTER
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
ASCENSION ORTHOPEDICS, INC.
11101 metric blvd
austin TX 78758
Manufacturer (Section G)
ASCENSION ORTHOPEDICS, INC.
11101 metric blvd
austin TX 78758
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14655116
Report Number3002788818-2022-00036
Device Sequence Number5109493
Product Code LXH
UDI-Device Identifier10381780237051
UDI-Public10381780237051
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberHDL096006501
Device Catalogue NumberHDL-0960-065-01
Device Lot Number27242206
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/24/2022
Supplement Date Received by Manufacturer07/26/2022
11/02/2023
Initial Report FDA Received Date06/10/2022
Supplement Report FDA Received Date07/28/2022
11/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-