1 of 2 reports.Other mfg report number: 3013886523-2022-00261.A physician reported a certas valve (id 828804 ) was implanted due to sah (subarachnoid hemorrhage) via v-p shunt on october 12, 2020 with unknown setting.It was used with peritoneal catheter 823045.Since obstruction was suspected, the valve and the peritoneal catheter were removed and replaced on (b)(6) 2022.
|
Updated fields: d4, d9, g3, g6, h2, h3, h6, h10 certas valve was returned for evaluation.Failure analysis- the position of the cam when valve was received was at setting 3.The valve was visually inspected; no defects were noted.The valve passed the test for programming, flushing, reflux, siphon guard and pressure.The valve was leak tested and it leaked from the needle holes in the needle chamber.No root cause could be determined as the technician could not confirm any problem with the valve at the time of investigation.The possible root cause for the issue reported by the customer could be due to biological debris and protein build up interfering with the valve mechanism, but at the time of investigation no occlusion issues were noted.
|