ETHICON INC. BLAKE CARDIO CONNECTOR 2:1; CATHETER, IRRIGATION Back to Search Results
Model Number BCC2
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Additional information: the product was placed into intracardiac and anterior mediastinum and connected to a low-pressure continuous suction equipment.At that time, a y-type connector was connected.After the surgery, when an experiment was performed to flow water using the product, water flowed out from the drain rupture part.After the leakage occurred, the product was replaced by the doctor.It was reported that no health hazard occurred to the patient.The doctor commented that a metal part of a milking roller was the cause of the breakage.Additional information has been requested and received.What is the lot number for each of the involved products? no further information is available.Please clarify the number of products involved in the event and how many to be returned? qty of product involved and to be returned of bcc2 is 1.Qty of product involved and to be returned of 2232 is 2.Device return status we regularly contact with sales rep about the device returning.No further information will be provided.Additional information has been requested and received.Attempts have been made to obtain the device, not received to date.If further details are received at a later date a supplemental medwatch will be sent.Response received stating "product was replaced by the doctor" was surgical intervention performed to place 2 new drains? please provide details and dates.Note: events reported on mw# 2210968-2022-04474, mw# 2210968-2022-04475.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported a patient underwent an aortic valve replacement on (b)(6) 2022 and a connector was used.After the surgery, in the icu, the surgeon confirmed that air leakage occurred at the connection part.The connection between the connector and the drain was held together by something like a zip tie.A drain was found broken in the area in contact with the tip of the connector.The doctor commented that a metal part of a milking roller was the cause of the breakage.There were no adverse consequences to the patient.Additional information has been requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Additional information has been requested and received.Attempts have been made to obtain the device, however not received to date.If further details are received at a later date a supplemental medwatch will be sent.Response received stating product was replaced by the doctor was surgical intervention performed to place 2 new drains? no further information is available.Please provide details and dates.Event date was 2022/05/17.No further information will be provided.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Additional information has been requested and received.Attempts to obtain the device have been made.If further details are received at a later date a supplemental medwatch will be sent.Date of index surgical procedure? 2022/05/17.Was the date the drain observed as broken 5/17/2022? yes.Was a 2nd procedure performed to place 2 new drains via surgical intervention? no further information is available.If yes, what was the date of drain replacement procedure? who monitored the drainage and how often? no further information is available.What symptoms did the patient experience following the index surgical procedure? onset date? no further information is available.Other relevant patient history/concomitant medications? no further information is available.What is the physician¿s opinion as to the etiology of or contributing factors to this event? the surgeon said that the breaking had been caused by a milking roller.What is the patient's current status? no further information is available.Product lot number? no further information is available.Will product be returned, please provide the return date and tracking information? the device was shipped.No further information will be provided." this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Additional information: d9.H3 evaluation: the product was returned for evaluation.One used sample of y type cardio connector received for evaluation both ports of connector were clamped with piece of drain at each side with the help of a zip ties.No manufacturing defect was observed.Improper handling and improper usage at user end could not be ruled out.Further investigation of the sample was not possible.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLAKE CARDIO CONNECTOR 2:1
Common Device Name
CATHETER, IRRIGATION
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
DEGANIA INDIA
251, sector-6, imt manesar
gurugram
IN  
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
MDR Report Key14656254
Report Number2210968-2022-04473
Device Sequence Number17360476
Product Code GBX
UDI-Device Identifier10705031023321
UDI-Public10705031023321
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
CL I EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 08/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBCC2
Device Catalogue NumberBCC2
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/24/2022
Supplement Date Received by Manufacturer06/14/2022
07/06/2022
08/10/2022
Initial Report FDA Received Date06/10/2022
Supplement Report FDA Received Date07/06/2022
07/26/2022
08/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age69 YR
Patient SexMale
-
-