| Model Number |
180111 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
|
| |
|
Event or Problem Description
|
|
It was reported that a peritoneal dialysis (pd) patient experienced a fluid leak in the pump module area of the liberty select cycler after ending their pd treatment.It is unknown at which point in therapy the leak may have begun.The cause of the leak is unknown.Upon follow up, it was confirmed that the patient was able to complete peritoneal dialysis treatment.The patient did not develop any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The cassette used for treatment was discarded and is not available for return for physical evaluation by the manufacturer.Additional information was requested however to date has not been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d.9, h.3 corrected information: g.2, g.4 plant investigation: no parts were returned to the manufacturer for physical evaluation.A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The investigation into the cause of the reported problem was not able to be confirmed.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
|
| |
|
Event or Problem Description
|
|
It was reported that a peritoneal dialysis (pd) patient experienced a fluid leak in the pump module area of the liberty select cycler after ending their pd treatment.It is unknown at which point in therapy the leak may have begun.The cause of the leak is unknown.Upon follow up, it was confirmed that the patient was able to complete peritoneal dialysis treatment.The patient did not develop any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The cassette used for treatment was discarded and is not available for return for physical evaluation by the manufacturer.Additional information was requested however to date has not been provided.
|
| |
|