CONCORD MANUFACTURING UL LIBERTY CYCLER ASSEMBLY(NON-VALUATED); SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Model Number 180111
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event or Problem Description
It was reported that a peritoneal dialysis (pd) patient experienced a fluid leak in the pump module area of the liberty select cycler after ending their pd treatment.It is unknown at which point in therapy the leak may have begun.The cause of the leak is unknown.Upon follow up, it was confirmed that the patient was able to complete peritoneal dialysis treatment.The patient did not develop any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The cassette used for treatment was discarded and is not available for return for physical evaluation by the manufacturer.Additional information was requested however to date has not been provided.
 
Additional Manufacturer Narrative
Additional information: d.9, h.3 corrected information: g.2, g.4 plant investigation: no parts were returned to the manufacturer for physical evaluation.A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The investigation into the cause of the reported problem was not able to be confirmed.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
 
Event or Problem Description
It was reported that a peritoneal dialysis (pd) patient experienced a fluid leak in the pump module area of the liberty select cycler after ending their pd treatment.It is unknown at which point in therapy the leak may have begun.The cause of the leak is unknown.Upon follow up, it was confirmed that the patient was able to complete peritoneal dialysis treatment.The patient did not develop any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The cassette used for treatment was discarded and is not available for return for physical evaluation by the manufacturer.Additional information was requested however to date has not been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UL LIBERTY CYCLER ASSEMBLY(NON-VALUATED)
Common Device Name
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key14657997
Report Number0002937457-2022-00937
Device Sequence Number3914721
Product Code FKX
UDI-Device Identifier00840861100972
UDI-Public00840861100972
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K181108
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number180111
Device Catalogue NumberRTLR180111
Was Device Available for Evaluation? No
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/29/2022
Supplement Date Received by Manufacturer06/22/2022
Initial Report FDA Received Date06/10/2022
Supplement Report FDA Received Date06/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/19/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DELFLEX PD FLUID; DELFLEX PD FLUID; LIBERTY CYCLER SET ; LIBERTY CYCLER SET 
Patient SexFemale
-
-