STRYKER SPINE-US XIA 3 VITALIUM ROD WITH HEX 6.0MM X 600MM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Model Number 48232601
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/11/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a xia 3 vitalium rod fractured post-operatively.The patient was revised and the rod was explanted.
 
Additional Manufacturer Narrative
Hospital retained the device.
 
Additional Manufacturer Narrative
Visual, dimensional, material and functional analysis could not be performed as the device was not returned.Device history review could not be performed as a valid lot number was not provided and could not be obtained.Complaint history records were reviewed for this catalog number, no adverse trends were identified.It was reported via follow up communication that both rods fractured but cannot be returned.Lots #'s could not be provided.Construct was from t4-t10 and the rods fractured at t8.Patient did not experience any trauma or accident but only achieved partial fusion.X-ray provided was reviewed and confirms that both rods fractured just cranial to t8.A corpectomy cage is implanted in the t7 vertebral body space with a cross connector spanning t7 also.Screw placement and rod overhand appears sufficient at all levels.No abnormalities identified in x-ray.The device surgical technique guide was reviewed, the following was found to relevant: adverse effects ¿ while the expected life of spinal implant components is difficult to estimate, it is finite.These components are made of foreign materials which are placed within the body for the potential fusion of the spine and reduction of pain.However, due to the many biological, mechanical and physicochemical factors which affect these devices but cannot be evaluated in vivo, the components cannot be expected to indefinitely withstand the activity level and loads of normal healthy bone.¿ bending, disassembly or fracture of any or all implant components.¿ fatigue fracture of spinal fixation devices, including screws and rods, has occurred.¿ delayed union or nonunion: internal fixation appliances are load sharing devices which are used to obtain alignment until normal healing occurs.In the event that healing is delayed, does not occur, or failure to immobilize the delayed/nonunion results, the implant will be subject to excessive and repeated stresses which can eventually cause loosening, bending or fatigue fracture.The degree or success of union, loads produced by weight bearing, and activity levels will, among other conditions, dictate the longevity of the implant.If a nonunion develops or if the implants loosen, bend or break, the device(s) should be revised or removed immediately before serious injury occurs.The most likely cause of the reported event was determined to be the reported lack of complete fusion.Failure to fuse over 3 years means the implant(s) will be subject to excessive and repeated stresses which can eventually cause loosening, bending or fatigue fracture.
 
Event or Problem Description
It was reported that two xia 3 vitalium rod fractured post-operatively.The patient was revised and the rod was explanted.This report captures the first rod.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIA 3 VITALIUM ROD WITH HEX 6.0MM X 600MM
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
allendale NJ 07401
Manufacturer (Section G)
STRYKER SPINE-SWITZERLAND
le crêt-du-locle 10 a
-
la chaux-de-fonds 2300
SZ   2300
Manufacturer Contact
rita karan
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key14658017
Report Number3005525032-2022-00014
Device Sequence Number4004834
Product Code NKB
UDI-Device Identifier04546540608642
UDI-Public04546540608642
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K113666
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number48232601
Device Catalogue Number48232601
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/11/2022
Supplement Date Received by Manufacturer08/05/2022
Initial Report FDA Received Date06/10/2022
Supplement Report FDA Received Date08/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient SexMale
Patient Weight90 KG
-
-