• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SALTER LABS SALTER LABS; INFUSOR,REUSABLE,INFUSEIT 500 ML W/STOPCOCK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SALTER LABS SALTER LABS; INFUSOR,REUSABLE,INFUSEIT 500 ML W/STOPCOCK Back to Search Results
Model Number ZIT-520-5
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 04/22/2022
Type of Reportable Event Malfunction
Event or Problem Description
The gauge disconnected from the tubing.
 
Additional Manufacturer Narrative
Reportable as disconnection happened during use delaying patient therapy.Capa-00463 was initiated for similar complaint.Manufacturing date for lot 20210927 pre-dates the implementation activities of capa-00463.Customer was notified that this complaint, along with several others submitted by the same customer, are tied to capa-00463 and an update would be provided once capa actions are implemented.Customer has not requested additional follow up at this time (reference (b)(4).
 
Event or Problem Description
The gauge disconnected from the tubing.
 
Additional Manufacturer Narrative
Reportable as disconnection happened during use delaying patient therapy.(b)(4) was initiated for similar complaint.Manufacturing date for lot 20210927 pre-dates the implementation activities of (b)(4).Customer was notified that this complaint, along with several others submitted by the same customer, are tied to capa-00463 and an update would be provided once capa actions are implemented.Customer has not requested additional follow up at this time (reference complaint-07296) summary: capa-00463 created for this reported failure mode.Manufacturing lot pre-dates capa corrective action implementation.Ra: pressure leaking caused by inadequate or defective materials including gauges is identified in the risk analysis at a 6 (major).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SALTER LABS
Common Device Name
INFUSOR,REUSABLE,INFUSEIT 500 ML W/STOPCOCK
Manufacturer (Section D)
SALTER LABS
30 spur drive
el paso TX 79906
Manufacturer (Section G)
SALTER LABS
30 spur drive
el paso TX 79906
Manufacturer Contact
melissa brickley
2710 northridge dr nw suite a
grand rapids, MI 49544
6162598415
MDR Report Key14658263
Report Number3000219639-2022-00026
Device Sequence Number17303097
Product Code KZD
UDI-Device Identifier10607411916912
UDI-Public10607411916912
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberZIT-520-5
Device Catalogue NumberZIT-520-5
Device Lot Number20210927
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/12/2022
Supplement Date Received by Manufacturer05/12/2022
Initial Report FDA Received Date06/10/2022
Supplement Report FDA Received Date06/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-