• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC WAYNE PNEUMOTHORAX TRAY; JOL CATHETER AND TIP, SUCTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC WAYNE PNEUMOTHORAX TRAY; JOL CATHETER AND TIP, SUCTION Back to Search Results
Model Number N/A
Medical Device Problem Codes Improper or Incorrect Procedure or Method (2017); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pneumothorax (2012)
Date of Event 04/13/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient required a wayne pneumothorax tray for a chest tube placement.After placement, it was discovered the user had inserted the chest drain valve backwards, not allowing air to exit the chest/pleural space.This resulted in worsening of the patient's pneumothorax.It is not known at this time how much time elapsed from placement until discovery and what specific interventions were performed.The device was explanted on (b)(6) 2022.No other adverse effects were reported for this incident.Additional information regarding the event and patient outcome has been requested but is currently unavailable.
 
Additional Manufacturer Narrative
Occupation: risk manager.Pma/510(k) #: exempt.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Correction: h6 - annex a.Investigation ¿ evaluation.It was reported by presbyterian health service corp (united states) that on 13apr2022 the chest drain valve in a wayne pneumothorax tray (rpn: c-utpty-1400-wayne-112497-imh; lot#: 14121501) was placed backwards.As a result, the patient¿s pneumothorax worsened.The device was explanted on (b)(6) 2022.No other adverse effects were reported due to this occurrence.Reviews of the complaint history, device history record (dhr), quality control, specifications, and instructions for use (ifu), were conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the dhr for the reported complaint device lot 14121501 and related sub-assembly lots revealed no recorded non-conformances.A database search did not identify any other events associated with the reported device lot.Based on the available information, cook has concluded that the device was manufactured to specification and that there is no evidence suggesting nonconforming product exists either in house or in field.Cook also reviewed product labeling.The ifu (c_t_waynemod_rev5) supplied with the complaint lot states the following in consideration of the reported failure mode: ¿instructions for use: 9.Remove the wire guide and catheter obturator.Attach connecting tube with stopcock, and cook chest drain valve.Attach cook chest drain valve in direction indicated by arrow on valve.Note: chest drain valve may be obviated if catheter is to be connected to a water seal suction apparatus or similar mechanical suction device.Do not connect catheter directly to wall suction.How supplied: upon removal from package, inspect the product to ensure no damage has occurred.¿ this product is supplied with an additional label attached to the chest drain valve indicating the correct direction to attach the connecting tube.Based on the available information, no device return, and the results of the investigation, the root cause for this event is user error.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WAYNE PNEUMOTHORAX TRAY
Common Device Name
JOL CATHETER AND TIP, SUCTION
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key14658820
Report Number1820334-2022-01043
Device Sequence Number3739827
Product Code JOL
UDI-Device Identifier00827002565371
UDI-Public(01)00827002565371(17)220701(10)14121501
Combination Product (Y/N)Y
Initial Reporter StateNC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/01/2022
Device Model NumberN/A
Device Catalogue NumberC-UTPTY-1400-WAYNE-112497-IMH
Device Lot Number14121501
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Fifteen-Day
Initial Date Received by Manufacturer 05/25/2022
Supplement Date Received by Manufacturer11/28/2022
Initial Report FDA Received Date06/10/2022
Supplement Report FDA Received Date12/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient SexUnknown
-
-