Model Number ACROBAT-I POSITIONER
Medical Device Problem Code Positioning Problem (3009)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
The hospital reported that during a coronary artery bypass procedure the acrobat-i positioner didn't engage when turning the knob and it was in the case.It didn't cause any effects or issues but they had to straighten the flexlink arm and re-position for it to tighten.No patient effects and they were able to complete the case with the device.
 
Additional Manufacturer Narrative
Trackwise id (b)(4).Since the device is not available to be returned to us, a technical evaluation cannot be performed.Per our standard sop's, all events are tracked and trended to determine whether or not any trends develop.Device discarded.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Trackwise # (b)(4).Updated section: b4, e3, g4, g7, h2, h3, h6, h10, h11 corrected section: udi # correction the device was not returned to maquet cardiac surgery for investigation, however a video was provided by the account.A video evaluation was conducted.The video was of a clinical procedure.The device arm was not visible in the video.A mechanical evaluation was conducted.The knob of the device was tightened, however we are unable to determine if the tightening of the knob had any effect on the flexarm as it was unable to been seen during the procedure in the video.Based off the video inspection, the reported failure "positioning problem" was not confirmed.The lot history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACROBAT-I POSITIONER
Common Device Name
STABILIZER, HEART
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key14660531
Report Number2242352-2022-00511
Device Sequence Number5937795
Product Code MWS
UDI-Device Identifier00607567500006
UDI-Public00607567500006
Combination Product (Y/N)N
Initial Reporter StateGA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/19/2023
Device Model NumberACROBAT-I POSITIONER
Device Catalogue NumberC-XP-5000Z
Device Lot Number3000183798
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/18/2022
Supplement Date Received by Manufacturer08/16/2022
Initial Report FDA Received Date06/10/2022
Supplement Report FDA Received Date08/16/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age72 YR
Patient SexFemale
-
-