PENUMBRA, INC. NEURON 6F 070 DELIVERY CATHETER; DQY Back to Search Results
Catalog Number PND6F0701058M
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/27/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
 
Event or Problem Description
The patient was undergoing a medical procedure in the left posterior inferior cerebellar artery (pica) using a neuron 6f 070 delivery catheter (neuron 070) and a non-penumbra short sheath.During the procedure, after introducing the neuron 070 via the short sheath, the physician noticed that the proximal end close to the hub of the neuron 070 was broken.Subsequently, it was noticed that some blood and saline came out of the broken part of the neuron 070.Therefore, the neuron 070 was removed.The procedure was completed using a new neuron 070 and the same short sheath.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Evaluation of the returned neuron 070 confirmed that the catheter was fractured.If the neuron 070 is manipulated at an extreme angle during use, damage such as a kink and subsequent fracture may occur.Further evaluation revealed an ovalization on the catheter shaft.This damage was incidental to the reported complaint and the root cause could not be determined.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEURON 6F 070 DELIVERY CATHETER
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key14661238
Report Number3005168196-2022-00277
Device Sequence Number20317740
Product Code DQY
UDI-Device Identifier00814548010052
UDI-Public814548010052
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K082290
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/23/2022
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/28/2024
Device Catalogue NumberPND6F0701058M
Device Lot NumberF103568
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/27/2022
Supplement Date Received by Manufacturer09/07/2022
Initial Report FDA Received Date06/10/2022
Supplement Report FDA Received Date09/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age57 YR
Patient SexFemale
Date Report Sent to FDA01/01/2005
-
-