H3, h6: the associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The post of the device is broken off, rendering the device inoperative.The post was returned.The surfaces of the device are scratched.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.The clinical/medical evaluation concluded that per complaint details, a revision was performed due to a broken insert.It was communicated that all broken pieces were removed prior to implanting the new poly and the requested medical documentation was not available.The current health status of patient remains unknown.Without the requested clinical documentation, the root cause of the reported event could not be fully assessed.The patient impact beyond the reported insert breakage and revision could not be determined.No further medical assessment could be rendered at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file and instructions for use document revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.As the device broke and it can no longer fit its purpose, the contribution of the device to the reported event could be corroborated.Some potential probable causes for this event could include but not limited to surgical technique or excessive forces applied to implant.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
|