• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LGN XLPE PS INSERT SZ 7-8 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LGN XLPE PS INSERT SZ 7-8 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 71453258
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/27/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after tka a a revision surgery was performed on (b)(6) 2022 due to a lgn xlpe ps insert sz 7-8 11mm broke inside the patient.All the pieces were removed prior to implanting a new poly.It is unknown if there was any surgical delay.No further information is available.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
H3, h6: the associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The post of the device is broken off, rendering the device inoperative.The post was returned.The surfaces of the device are scratched.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.The clinical/medical evaluation concluded that per complaint details, a revision was performed due to a broken insert.It was communicated that all broken pieces were removed prior to implanting the new poly and the requested medical documentation was not available.The current health status of patient remains unknown.Without the requested clinical documentation, the root cause of the reported event could not be fully assessed.The patient impact beyond the reported insert breakage and revision could not be determined.No further medical assessment could be rendered at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file and instructions for use document revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.As the device broke and it can no longer fit its purpose, the contribution of the device to the reported event could be corroborated.Some potential probable causes for this event could include but not limited to surgical technique or excessive forces applied to implant.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN XLPE PS INSERT SZ 7-8 11MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14670162
Report Number1020279-2022-02906
Device Sequence Number17436458
Product Code JWH
UDI-Device Identifier00885556255797
UDI-Public00885556255797
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K071071
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/08/2021
Device Model Number71453258
Device Catalogue Number71453258
Device Lot Number11BM07525
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/06/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/27/2022
Supplement Date Received by Manufacturer08/08/2022
Initial Report FDA Received Date06/11/2022
Supplement Report FDA Received Date08/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/11/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-