Model Number 74122542
Medical Device Problem Code Biocompatibility (2886)
Health Effect - Clinical Code Metal Related Pathology (4530)
Date of Event 09/20/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: case (b)(4).
 
Event or Problem Description
Us legal.It was reported that, after a bhr-tha construct had been implanted on the patient¿s left hip on an unspecified date, the patient developed osteolysis around her pelvis and loosening of the acetabular component following a metallosis diagnosis.A revision surgery was performed on (b)(6) 2021 to treat this adverse event.During this procedure, the posterior hip capsule had staining consistent with a gray metallosis, and the joint presented some grayish fluid and significant gray and black staining to the soft tissue.Some osteolysis along the posterior femoral shaft in the metaphyseal region was also noted.Upon removing the bhr modular femoral head and cup, the surgeon faced several difficulties whilst fixing a 56mm acetabular cup and was not satisfied with the outcome, as a properly fixation could had been compromised mainly due to the deficient condition of the medial, anterior and posterior wall of the acetabulum.Because of this, he decided to place the new implants and close the incision to further plan a caged reconstruction of the acetabulum.Patient¿s current health status is unknown.
 
Additional Manufacturer Narrative
H3, h6: it was reported that left hip revision surgery was performed due to osteolysis and metallosis.As of today, the implanted devices, all of which were used in treatment have not been returned for evaluation.Without a definitive batch number, a complete review of the historical complaints data cannot be performed for the device.A review of historical complaints data was performed using the part numbers and the reported failure modes to evaluate patterns of repeated failures or defects.Similar complaints were identified for the hemi head.However, as the device is no longer sold, no action is to be taken.In the absence of the actual device, the production records were reviewed for the devices reportedly involved in this incident.All released devices involved met manufacturing specifications upon release for distribution.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.A review of historic escalation actions related to the products and similar complaint events was performed.Following the review, prior applicable escalation actions were identified and confirmed to reduce associated risks as far as possible.No further escalation actions required.The available medical documents were reviewed.The clinical information provided of the gray and black staining to the soft tissue, grayish fluid, osteolysis, loose acetabular component, and bone deficiency may be consistent with a reaction to metal debris.However, the source and the root cause cannot be determined with the available documentation.It cannot be concluded that the reported clinical findings are associated with the implant failure.Based on the available information we can confirm the reported complaint, however without further information our investigation remains inconclusive, and a definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement, loosening of components may increase production of wear particles and accelerate damage to the bone.Should the device or additional information be received, the complaint will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.
 
Additional Manufacturer Narrative
H3, h6: it was reported that left hip revision surgery was performed.As of today, the implanted devices, all of which were used in treatment have not been returned for evaluation.Without definitive lot numbers a complete complaint history review cannot be performed for the devices involved.Should further details become available at a later date then the task will be reopened and completed.A review of the historical complaints data for the 74121342 modular head 42 mm -4 (short) and the acetabular cup was performed using part numbers and the reported failure modes to evaluate patterns of repeated failures or defects in a timeframe prior 12 months as of the complaint aware/reopen date.No other similar complaints were identified for the part numbers and the reported/related failure modes.However, as the devices are no longer sold, no action is to be taken.Without definitive lot numbers a complete complaint history review cannot be performed for the devices involved.Should further details become available at a later date then the task will be reopened and completed.As no device batch numbers were provided for investigation, manufacturing record review, device labelling / ifu review and historical review of the escalation actions could not be performed.If more information is received, this investigation will be reopened.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.The available medical documents were reviewed.The clinical information provided of the gray and black staining to the soft tissue, grayish fluid, osteolysis, loose acetabular component, and bone deficiency may be consistent with a reaction to metal debris.However, the source and the root cause cannot be determined with the available documentation.It cannot be concluded that the reported clinical findings are associated with the implant failure.Based on the available information we can confirm the reported complaint, however without further information our investigation remains inconclusive, and a definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement, loosening of components may increase production of wear particles and accelerate damage to the bone.Should the devices or additional information be received, the complaint will be re-opened.Based on this investigation, the need for corrective and preventative actions is not indicated.
 
Additional Manufacturer Narrative
H3, h6.It was reported that left hip revision surgery was performed due to osteolysis and metallosis.As of today, the implanted devices, all of which were used in treatment have not been returned for evaluation.Without a definitive batch number, a complete review of the historical complaints data cannot be performed for the device.A review of historical complaints data was performed using the part numbers and the reported failure modes to evaluate patterns of repeated failures or defects.Similar complaints were identified for the hemi head.However, as the device is no longer sold, no action is to be taken.As no device batch numbers were provided for investigation, a manufacturing record review could not be performed.If more information is received, this investigation will be reopened.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.A review of historic escalation actions related to the products and similar complaint events was performed.Following the review, prior applicable escalation actions were identified and confirmed to reduce associated risks as far as possible.No further escalation actions required.The available medical documents were reviewed.The clinical information provided of the gray and black staining to the soft tissue, grayish fluid, osteolysis, loose acetabular component, and bone deficiency may be consistent with a reaction to metal debris.However, the source and the root cause cannot be determined with the available documentation.It cannot be concluded that the reported clinical findings are associated with the implant failure.Based on the available information we can confirm the reported complaint, however without further information our investigation remains inconclusive, and a definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement, loosening of components may increase production of wear particles and accelerate damage to the bone.Should the device or additional information be received, the complaint will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMI HEAD 42MM
Common Device Name
PROSTHESIS, HIP, HEMI-, FEMORAL, METAL
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK   CV31 3HL
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14670396
Report Number3005975929-2022-00395
Device Sequence Number20475121
Product Code KWL
UDI-Device Identifier00885556070710
UDI-Public00885556070710
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 10/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number74122542
Device Catalogue Number74122542
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/25/2022
Supplement Date Received by Manufacturer09/06/2022
09/29/2022
09/26/2022
10/20/2022
Initial Report FDA Received Date06/11/2022
Supplement Report FDA Received Date09/13/2022
09/30/2022
10/19/2022
10/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
3005975929-3/10/15-001
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age50 YR