| Model Number |
1012447-12 |
| Medical Device Problem Codes |
Material Separation (1562); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported difficulty advancing the device and separation appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat the distal right coronary artery (rca).The 2.5x12mm trek rx balloon dilatation catheter (bdc) had resistance with the anatomy during advancement and the proximal shaft snapped in 2 separate pieces.There was no resistance during withdrawal and the separated portion was simply withdrawn.Another nc trek rx was used to complete the procedure.There was no adverse patient effect and there was no clinically significant delay in the procedure.No additional information was provided.
|
| |
|