SMITH & NEPHEW, INC. REDAPT TRIAL BODY SZ 12-15; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER
|
Back to Search Results |
|
| Model Number |
71354631 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/27/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that, during a thr procedure, while trying to extract from the patient the proximal redapt trial body sz 12-15 broke.All pieces of the mechanism were found and extracted via c-arm.Surgery was resumed, with a less than or equal to 30 minutes delay, with the same device.Patient was not harmed.No further complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference case (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device's wave spring was found to be broken and removed from the devices assembly, rendering the device inoperative.The device shows signs of extensive use.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.As the device broke and it can no longer fit its purpose, the contribution of the device to the reported event could be corroborated.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|