• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL AMPLATZER MUSCULAR VSD OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL AMPLATZER MUSCULAR VSD OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER Back to Search Results
Model Number VSD MUSCULAR OCCLUDER
Medical Device Problem Codes Migration or Expulsion of Device (1395); Off-Label Use (1494)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/13/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported a 14mm amplatzer muscular vsd occluder was chosen for implant on (b)(6) 2022.The device was sized by fluoroscopy.During procedure, post deployment, the device immediately embolized into the pulmonary artery.The patient remained hemodynamically stable/ a snare was used for retrieval and the device was removed.A replacement device was not used, the procedure was aborted.The patient's status was unknown.No additional information was provided.
 
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
Subsequent to the previously filed report, additional information was received that the cause of the embolization of the 14mm occluder was believed to be due to the device not fitting well and as well as the patient's anatomy.It was also reported that both devices were attempting to be placed in a patent ductus arteriosus (pda), which was off-label use.
 
Additional Manufacturer Narrative
An event of device embolization into the pulmonary artery was reported.A returned device assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.Please note that per the instructions for use, "indications and usage: the amplatzer¿ muscular vsd occluder is indicated for use in patients with a complex ventricular septal defect (vsd) of significant size to warrant closure (large volume left-to-right shunt, pulmonary hypertension, and/or clinical symptoms of congestive heart failure) who are considered to be at high risk for standard transatrial or transarterial surgical closure based on anatomical conditions and/or based on overall medical condition."na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER MUSCULAR VSD OCCLUDER
Common Device Name
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL REG# 2135147
5050 nathan ln n
plymouth MN 55442
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key14671998
Report Number2135147-2022-00373
Device Sequence Number17436462
Product Code MLV
UDI-Device Identifier00811806011837
UDI-Public00811806011837
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P040040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVSD MUSCULAR OCCLUDER
Device Catalogue Number9-VSD-MUSC-014
Device Lot Number8259441
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/13/2022
Supplement Date Received by Manufacturer09/01/2022
Initial Report FDA Received Date06/12/2022
Supplement Report FDA Received Date09/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age9 YR
Patient SexMale
-
-