• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS AG CS/CSL/GERADSCHAFT-PLUS EXTRACT.SCREW M6; OSTEOTOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS AG CS/CSL/GERADSCHAFT-PLUS EXTRACT.SCREW M6; OSTEOTOME Back to Search Results
Model Number 75002165
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Smith & nephew is submitting this report pursuant to the provisions of 21 c.F.R.Part 803.This report may be based upon information which smith & nephew has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, smith & nephew, or its employees, that the report constitutes an admission that the device, smith & nephew or its employees caused or contributed to the potential event described in this report.(b)(4).
 
Event or Problem Description
It was reported that during thr, the cs/csl/geradschaft-plus extract.Screw m6 broke while inside the patient.The surgery finished with a competitor's device.No injury or significant reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
H10: additional information received by the manufacturer identified that this event should be re-evaluated for mdr reporting.The new information received confirmed that the tip of the cs/csl/geradschaft-plus extract.Screw m6 is bent and that no piece is broken off.The reassessment determined that the issue does not meet the threshold for reporting and therefore, it is a non-reportable event.If further details are provided, our files will be updated accordingly, and a further report submitted outlining both the event details and our investigations performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS/CSL/GERADSCHAFT-PLUS EXTRACT.SCREW M6
Common Device Name
OSTEOTOME
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ   CH-5000
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14672459
Report Number9613369-2022-00283
Device Sequence Number4856190
Product Code HWM
UDI-Device Identifier07611996056563
UDI-Public07611996056563
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number75002165
Device Catalogue Number75002165
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/02/2022
Supplement Date Received by Manufacturer06/02/2022
Initial Report FDA Received Date06/12/2022
Supplement Report FDA Received Date06/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-