Model Number 6570-0-032
Medical Device Problem Code Inaccurate Information (4051)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/08/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a surgeon noticed the actual implanted head was not -4mm, it was +4mm head.
 
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.This is a product mix issue, so the device information located on the outer box label is being reported in section d and the information for the device that was actually inside the package is : cat# 6570-0-232; delta v-40 ceramic head 32/+4; lot# 89546202.Product field action has been initiated.Investigation ongoing.
 
Additional Manufacturer Narrative
Reported event: an event regarding product mix involving a ceramic head was reported.The event is in the scope of nc and therefore the event is confirmed.Method & results: product evaluation and results: not performed as the device remained implanted.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been 8 other similar events for the lot referenced.Conclusions: the reported product mix was confirmed.Nc was raised to further address this event.
 
Event or Problem Description
It was reported that a surgeon noticed the actual implanted head was not -4mm, it was +4mm head.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTA V-40 CERAMIC HEAD 32/-4
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
marisol santiago
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key14672585
Report Number0002249697-2022-00818
Device Sequence Number4855482
Product Code LZO
UDI-Device Identifier04546540608475
UDI-Public04546540608475
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K052718
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 07/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6570-0-032
Device Catalogue Number6570-0-032
Device Lot Number89648802
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/17/2022
Supplement Date Received by Manufacturer07/19/2022
Initial Report FDA Received Date06/12/2022
Supplement Report FDA Received Date07/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
PFA 2902313
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-