BOSTON SCIENTIFIC CORPORATION EQUALIZER; CATHETER, PERCUTANEOUS Back to Search Results
Model Number 20280
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial reporter city: (b)(6).
 
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in abdominal aorta.A 33/7/2/100 equalizer balloon catheter was advanced for dilatation.However, during the initial inflation at 5 seconds, the balloon was ruptured.The device was removed without any problem and the procedure was completed with another of same device.There were no patient complications nor injuries reported and the patient condition was good.
 
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in abdominal aorta.A 33/7/2/100 equalizer occlusion balloon catheter was advanced for dilatation.However, during the initial inflation at 5 seconds, the balloon was ruptured.The device was removed without any problem and the procedure was completed with another of same device.There were no patient complications nor injuries reported and the patient condition was good.
 
Additional Manufacturer Narrative
E1-initial reporter city: (b)(6).Device evaluated by mfr.: unit returned with its original box, overall visual revision did not identify failures or evidence that could be lost due to decontamination process.The balloon was ruptured.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUALIZER
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road
cork
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key14672909
Report Number2134265-2022-06695
Device Sequence Number5397789
Product Code MJN
UDI-Device Identifier08714729402404
UDI-Public08714729402404
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K140273
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/09/2022
Device Model Number20280
Device Catalogue Number20280
Device Lot Number0026594650
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/06/2022
Supplement Date Received by Manufacturer07/13/2022
Initial Report FDA Received Date06/13/2022
Supplement Report FDA Received Date07/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/07/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-