| Model Number |
6465-6-018 |
| Medical Device Problem Codes |
Degraded (1153); Fracture (1260); Unstable (1667); Naturally Worn (2988)
|
| Health Effect - Clinical Code |
Joint Laxity (4526)
|
| Date of Event |
05/20/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
Revision (hmrs parts replacement after 12 years) according to bfarm special approval 92.16-5640-s-023/22.Update 25may22: after 12 years , the poly is worn out.There was a massive instability in the knee joint.Especially the stopper on the proximal tibia was very crushed.
|
| |
|
Event or Problem Description
|
|
Revision (hmrs parts replacement after 12 years) according to bfarm special approval 92.16-5640-s-023/22.Update 25may22.After 12 years , the poly is worn out.There was a massive instability in the knee joint.Especially the stopper on the proximal tibia was very crushed.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding crack/fracture, wear, and instability involving a hmrs bushing was reported.The crack/fracture and wear events were confirmed via evaluation of the returned device.Instability was not confirmed.Method & results: -product evaluation and results: visual inspection: the damage present on the hmrs bushing is consistent with delamination and material loss; which are common damage modes of uhmwpe.The bushing has fractured along the body of the articulating surface.Hackles are observed on the fractured surface are consistent with a fracture in overload.-clinician review: no medical records were received for review with a clinical consultant.-product history review: could not be performed as the device lot details were not provided.-complaint history review: could not be performed as the device lot details were not provided.Conclusions: it was reported that the patient was revised due to worn poly components and joint instability.The damage present on the hmrs bushing is consistent with delamination and material loss; which are common damage modes of uhmwpe.The bushing has fractured along the body of the articulating surface.Hackles are observed on the fractured surface are consistent with a fracture in overload.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Search Alerts/Recalls
|