STRYKER ORTHOPAEDICS-MAHWAH HMRS WEDGE - 7 X 18; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 6465-6-007
Medical Device Problem Codes Degraded (1153); Fracture (1260); Unstable (1667); Naturally Worn (2988)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 05/20/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
Revision (hmrs parts replacement after 12 years) according to bfarm special approval 92.16-5640-s-023/22.Update 25may22: after 12 years , the poly is worn out.There was a massive instability in the knee joint.Especially the stopper on the proximal tibia was very crushed.
 
Additional Manufacturer Narrative
Reported event: an event regarding crack/fracture, wear, and instability involving a hmrs wedge was reported.The crack/fracture and wear events were confirmed via evaluation of the returned device.Instability was not confirmed.Method & results: -product evaluation and results: visual inspection: the damage present on the hmrs wedge is consistent with delamination and material loss; which are common damage modes of uhmwpe.The wedge appears to have split open.Hackles are observed on the fractured surface are consistent with a fracture in overload.-clinician review: no medical records were received for review with a clinical consultant.-product history review: could not be performed as the device lot details were not provided.-complaint history review: could not be performed as the device lot details were not provided.Conclusions: it was reported that the patient was revised due to worn poly components and joint instability, a "crushed" stopper was also reported.The damage present on the hmrs wedge is consistent with delamination and material loss; which are common damage modes of uhmwpe.The wedge appears to have split open.Hackles are observed on the fractured surface are consistent with a fracture in overload.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Revision (hmrs parts replacement after 12 years) according to bfarm special approval 92.16-5640-s-023/22.Update 25may2022.After 12 years , the poly is worn out.There was a massive instability in the knee joint.Especially the stopper on the proximal tibia was very crushed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HMRS WEDGE - 7 X 18
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
EI   NA
Manufacturer Contact
diana rogers
210 centennial avenue
centennial park, elstree
borehamwood, NJ WD6 3-SJ
UK   WD6 3SJ
2018315000
MDR Report Key14674068
Report Number0002249697-2022-00823
Device Sequence Number17547200
Product Code KRO
UDI-Device Identifier07613327048933
UDI-Public07613327048933
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K122015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number6465-6-007
Device Catalogue Number64656007
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/20/2022
Supplement Date Received by Manufacturer11/11/2022
Initial Report FDA Received Date06/13/2022
Supplement Report FDA Received Date11/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age64 YR
Patient SexMale
-
-