WILLIAM COOK EUROPE ZENITH ALPHA¿ THORACIC ENDOVASCULAR GRAFT PROXIMAL TAPERED COMPONENTS; MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number G38191
Medical Device Problem Codes Mechanical Jam (2983); Difficult or Delayed Separation (4044)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/31/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturer ref# (b)(4).Pma/510(k): p140016.(b)(4).Investigation is still in progress.
 
Event or Problem Description
Description of event according to initial reporter: a 74 year old female patient underwent a tvar procedure in which the zenith alpha throacic endovascular graft proximal tapered components, g38191, was used.The proform wire did not detach from implant.The district manager and surgeon broke apart the handle to manually release the wire and graft.The graft had pulled back from intended spot of deployment as the physician had thought the graft had detached from the delivery system.At this time there is no follow up procedure being scheduled to place a graft in the area above the implanted graft, however, the surgeon stated at some point in the future another procedure will need to be done.
 
Event or Problem Description
Additional information (b)(6) 2022: patient anatomy was normal in area of difficulty.The sequence on how the trigger wires were released: surgeon rotated handle on alpha device after deployed to release trigger wires.Handle stopped rotating but wires did not fully release from the proform suture because device pulled back when trying to take dilator out.If it was released it wouldn¿t be attached to delivery system anymore.Stopped pulling dilator and removed wires manually by removing handle.Graft probably came back a little less than 1cm from where it was originally.
 
Additional Manufacturer Narrative
Manufacturer ref#(b)(4).Investigation is still in progress.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Additional Manufacturer Narrative
Manufacturer ref#(b)(4).Summary of investigational findings: a 74-year-old female patient underwent tevar (thoracic endovascular aortic repair) in which a zta-pt-40-36-167-w (complaint device) was used.The steel wire did not detach from implant despite the rotation handle has been rotated to stop.The district manager and surgeon broke apart the handle to manually release the wire and graft.The graft had pulled back from intended spot of deployment approximately 1 cm as the physician had thought the graft had detached from the delivery system.At this time there is no follow up procedure being scheduled to place a graft in the area above the implanted graft, however, the surgeon stated at some point in the future another procedure will need to be done.No device was returned, and no imaging was provided.Review of the device history record gave no indication of the device being produced out of specification.Based on the provided information it has not been possible to determine the cause of the reported event.Cook medical will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZENITH ALPHA¿ THORACIC ENDOVASCULAR GRAFT PROXIMAL TAPERED COMPONENTS
Common Device Name
MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
Manufacturer (Section G)
WILLIAM COOK EUROPE
sandet 6, dk-4632
bjaeverskov
Manufacturer Contact
lissi walmann
sandet 6, dk-4632
bjaeverskov 
MDR Report Key14674165
Report Number3002808486-2022-00818
Device Sequence Number14837318
Product Code MIH
UDI-Device Identifier10827002381916
UDI-Public(01)10827002381916(17)240528(10)E4114241
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/28/2024
Device Model NumberG38191
Device Catalogue NumberZTA-PT-40-36-167-W
Device Lot NumberE4114241
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/31/2022
Supplement Date Received by Manufacturer07/02/2022
11/25/2022
Initial Report FDA Received Date06/13/2022
Supplement Report FDA Received Date07/18/2022
11/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age74 YR
Patient SexFemale
-
-