Model Number SN60WF
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/16/2022
Type of Reportable Event Malfunction
Event or Problem Description
A non-healthcare professional reported that during an intraocular lens (iol) implant procedure, crack was found on iol optic after implanted in the patient's eye, then removed.The surgery was completed after a new replacement was used.Additional information was requested.
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The product was returned for analysis and the reported complaint was observed.Iol returned in two segments in the iol case.Solution is dried on both surfaces of the optic and haptics.Both haptics are intact.The optic is cracked/fractured and scratched/marked-rejectable with fibers & particulate adhered to the solution we are unable to determine the root cause for the reported complaint "crack on iol optic".The returned iol shows evidence of possible handling by the customer due to the presence of a dried solution on both surfaces of the optic and haptics.In addition to this, all iols are 100% cosmetically inspected as per approved manufacturing procedures and the observed lens damage would not meet our current release criteria.Based on these investigation findings, we are unable to verify if the iol contributed to the event.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ NATURAL SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14674208
Report Number9612169-2022-00264
Device Sequence Number4855505
Product Code HQL
UDI-Device Identifier00380655093283
UDI-Public00380655093283
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/24/2026
Device Model NumberSN60WF
Device Catalogue NumberSN60WF.245
Device Lot Number25139271
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/17/2022
Supplement Date Received by Manufacturer08/11/2022
Initial Report FDA Received Date06/13/2022
Supplement Report FDA Received Date09/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-