| Model Number |
SN60WF |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
05/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A non-healthcare professional reported that during an intraocular lens (iol) implant procedure, crack was found on iol optic after implanted in the patient's eye, then removed.The surgery was completed after a new replacement was used.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
A sample device was not returned for analysis.Product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned for analysis and the reported complaint was observed.Iol returned in two segments in the iol case.Solution is dried on both surfaces of the optic and haptics.Both haptics are intact.The optic is cracked/fractured and scratched/marked-rejectable with fibers & particulate adhered to the solution we are unable to determine the root cause for the reported complaint "crack on iol optic".The returned iol shows evidence of possible handling by the customer due to the presence of a dried solution on both surfaces of the optic and haptics.In addition to this, all iols are 100% cosmetically inspected as per approved manufacturing procedures and the observed lens damage would not meet our current release criteria.Based on these investigation findings, we are unable to verify if the iol contributed to the event.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|