• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL W/ACRYSERT C DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL W/ACRYSERT C DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR Back to Search Results
Model Number SN6CWS
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/09/2022
Type of Reportable Event Malfunction
Event or Problem Description
A non-healthcare professional reported that during an intraocular lens (iol) implant procedure, crack was found on iol optic after implanted in the patient's eye, then removed.The surgery was completed after a new replacement was used.Additional information was requested.
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
Additional information was requested and received.Stating, this lens was removed from the device.And placed into a cartridge to implant.
 
Additional Manufacturer Narrative
The company device with the explanted lens was returned in the opened tray, inside the cardboard sleeve in the opened carton.The lens door was closed.The lens stop has been replaced into the lens door.There appears to be inadequate amount of solution in the device.The solution was only observed, on the lens in the loading area, at the nozzle entry area, where one haptic tip was extended, and a small spot in the tip.The dried viscoelastic did not resemble the typical company viscoelastic patterns.It is unknown, what specific viscoelastic was used.The plunger has been retracted to mid-loading area.And was pressed against the lens edge.The optic was cut into pieces, typical of damage created to facilitate removal from the eye.The lens pieces had been replaced into the loading area, prior to return.The portions were atop one another.One haptic extended into the nozzle entry area.The optic edge has a broken area.The broken portion was returned.Another broken area of the optic was observed, near one haptic.The optic edge near this damage was split and cracked in two areas.The damage traveled toward the optic center.A slim, linear slightly opaque colored foreign material was observed, on the posterior of the optic.The nozzle was removed and cleaned for further evaluation.Top coat dye stain testing was conducted with acceptable results.A photo was provided to review of the lens, while inside the eye.The returned product has a linear strip of foreign material on the optic, that is similar in shape to the linear shadowed shape seen in the photo.There was no crack observed, in the photo.The linear shaped artifact was most likely interpreted as the complaint.Follow up information, indicated the lens was removed from the company device and loaded into a cartridge.This is not an approved delivery method per, the ifu.Company lens is a preloaded single use product.The cartridge information was not provided by the customer.Follow up was provided, that discovers viscoelastic was used.The root cause for the reported event was most likely, due to a failure to follow the ifu.Follow up was provided, that indicated the lens was removed from the preloaded company device and then loaded into a cartridge.This is not an approved delivery method for the company product.The company delivery system is a preloaded device manufactured for single use only.A photo of the lens inside the eye was made available.There was no crack observed, in the photo.A linear artifact that matches the artifact on the returned cut lens was observed.This was most likely interpreted as the reported complaint.The slightly opaque colored artifact was isolated and sent to particle lab for identification and analysis.The sample was visually and microscopically examined and the reported foreign material was observed.Microscopic examination shows clear material approximately 2.5 mm in length.The clear material was then isolated and analyzed using microscopic fourier transform infrared spectroscopy (micro ft-ir).Comparison of the material¿s generated ir spectra to a library of spectra finds the best match to be monarch cartridge.The observations of the device evaluation, the explanted lens evaluation, the dye stain testing results, the photo evaluation, and the lab analysis support the customer's information, that this lens was not delivered through the company device.The lab testing, indicated best match of the material to be company cartridge material.However, "best match" is not a 100% confirmation.The customer did not indicate, the brand or model of cartridge that was incorrectly used in this event.No further information has been provided for the file.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ SINGLEPIECE IOL W/ACRYSERT C DELIVERY SYSTEM
Common Device Name
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14674209
Report Number1119421-2022-01250
Device Sequence Number4855506
Product Code KYB
UDI-Device Identifier00380652249287
UDI-Public00380652249287
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/22/2026
Device Model NumberSN6CWS
Device Catalogue NumberSN6CWS.235
Device Lot Number15257422
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/17/2022
Supplement Date Received by Manufacturer08/07/2022
08/31/2022
Initial Report FDA Received Date06/13/2022
Supplement Report FDA Received Date08/29/2022
09/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DISCOVISC; MONARCH III INJECTOR; UNSPECIFIED CARTRIDGE; UNSPECIFIED OPHTHALMIC VISCOSURGICAL DEVICE
Patient SexUnknown
-
-