| Model Number |
OSTEOAUGER BONE GRAFT HARVESTER, 6MM |
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
05/20/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
On (b)(6) 2022, it was reported by a sales representative via email that while harvesting bone from the calcaneus of a patient using an abs-8000-06 8mm osteo-auger, patient experienced bone burning.Surgeon down sized to a 6mm osteo-auger and was able to complete the desired diameter along with saline irrigation to decrease the amount of bone burning.This was discovered during a procedure on (b)(6) 2022.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Search Alerts/Recalls
|