• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. VG 360 5N1 LK TIB TRL 63/67X14; INSTRUMENT, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. VG 360 5N1 LK TIB TRL 63/67X14; INSTRUMENT, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the while doing a revision of competitor product , the 12mm and 14mm 5-1 bearing trials were damaged during removal from the trialing phase of the operation.No parts fell in the patient, no patient impact.
 
Additional Manufacturer Narrative
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual inspection of the returned trials confirms a piece was fractured off the posterior of the distal surface and all pieces were not returned.The device also exhibits wear consistent with usage of device.Review of the device history records identified no related deviations or anomalies during manufacturing.Reported event is not related to a combination of products; therefore a compatibility review is not applicable.Medical records were not provided.The reported lot has been in the field for approximately ten years.It is unknown for how many times the device is being used.The root cause can be contributed to normal wear and tear of device from repeated use over time.No corrective actions, preventive actions, or field actions resulted after investigation of this event.This complaint can be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VG 360 5N1 LK TIB TRL 63/67X14
Common Device Name
INSTRUMENT, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key14674630
Report Number0001825034-2022-01392
Device Sequence Number20104615
Product Code JWH
UDI-Device Identifier00880304515741
UDI-Public(01)00880304515741(10)886820
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K093293
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number32-360524
Device Lot Number886820
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/19/2022
Supplement Date Received by Manufacturer06/28/2022
Initial Report FDA Received Date06/13/2022
Supplement Report FDA Received Date07/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/01/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient Age52 YR
Patient SexFemale
-
-