| Model Number |
101/860/085CZ |
| Medical Device Problem Code |
Disconnection (1171)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A sample was received to perform an investigation.A visual inspection of the returned device, found that the suction connector was disconnected from the suction tube.The reported problem was confirmed through visual inspection.This was determined to be a supply item issue.The root cause was unable to be determined.The affected component was shipped for a secondary investigation.Operator of device is unknown.No product information has been provided to date.
|
| |
|
Event or Problem Description
|
|
It was reported that when the customer performed a suction through the upper part of the cuff during use, the connector at the end of the suction line detached.No additional information was available.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).No problems or issues were identified during this device history record review.A product sample was received for evaluation.Visual and functional testing were performed.According to the photos provided, it was confirmed that the connector was detached from the tube in the suction line of the subassembly.The complaint was confirmed.The root cause of the reported issue was found to be the connector at the end of the suction line got detached.Actions were taken to mitigate the reported issue: customer complaint notification was performed to the production personnel as awareness to failure mode reported.
|
| |
|
Search Alerts/Recalls
|