MAKO SURGICAL CORP. BONE PIN, 3.2MM X 140MM, STERILE 2 PACK; STEREOTAXIC DEVICE, ROBOTICS Back to Search Results
Model Number 143140
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 05/16/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.Device not returned.
 
Event or Problem Description
The 3.2 x 140 pins used to place the femoral array, when being removed with a drill, one pin broke off in the femur.The pa removing pin did not catch that it was broken and patient was sent to pacu where a post op x-ray detected the half broken off.Patient was brought back and the remaining pin was removed.
 
Event or Problem Description
The 3.2 x 140 pins used to place the femoral array, when being removed with a drill, one pin broke off in the femur.The pa removing pin did not catch that it was broken and patient was sent to pacu where a post op x-ray detected the half broken off.Patient was brought back and the remaining pin was removed.
 
Additional Manufacturer Narrative
Reported event.An event regarding crack/fracture involving a mako bone pins was reported.The event was not confirmed because the product was not available for inspection.Method & results.Product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: a review of the device history records cannot be conducted as the lot/serial number was not reported.Complaint history review: a complaint history review cannot be conducted as the lot/serial number was not reported.Conclusions: the alleged failure mode was not confirmed because the product was not available for inspection.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.H3 other text : device not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BONE PIN, 3.2MM X 140MM, STERILE 2 PACK
Common Device Name
STEREOTAXIC DEVICE, ROBOTICS
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
perla zima
2555 davie road
fort lauderdale, FL 33317
9546280700
MDR Report Key14675020
Report Number3005985723-2022-00075
Device Sequence Number17495364
Product Code OLO
UDI-Device Identifier00848486018146
UDI-Public00848486018146
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number143140
Device Catalogue Number143140
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/16/2022
Supplement Date Received by Manufacturer12/15/2022
Initial Report FDA Received Date06/13/2022
Supplement Report FDA Received Date12/29/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other; Hospitalization;
Patient Age87 YR
Patient SexFemale
-
-