STRYKER NEUROVASCULAR CORK TRANSFORM SUPER COMPLIANT 4MM X 7MM - US; CATHETER, INTRAVASCULAR OCCLUDING, TEMPORARY Back to Search Results
Model Number M003SSC04070
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Perforation of Vessels (2135)
Date of Event 05/19/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during the anterior cerebral artery procedure, the balloon (subject device) ruptured inside the patient's anatomy.Also, the patient blood vessel was ruptured.No further information available about the procedure and patient condition.
 
Additional Manufacturer Narrative
The device is not available to the manufacturer.
 
Additional Manufacturer Narrative
Due to the automated mes system, there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device is not available; therefore, functional testing as well as visual testing cannot be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.It was reported that the subject balloon ruptured inside the patient.Additional information provided by the customer indicated that no anomalies were noted to the device prior to preparation, the device was prepared as per the dfu, 50% contrast was used during preparation and procedure, continuous flush was maintained throughout the procedure, and the patient anatomy was moderately tortuous.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause and the device was not returned, an assignable cause of undeterminable was assigned to this complaint.
 
Event or Problem Description
It was reported that during the anterior cerebral artery procedure, the balloon (subject device) ruptured inside the patient's anatomy.Also, the patient blood vessel was ruptured.No further information available about the procedure and patient condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANSFORM SUPER COMPLIANT 4MM X 7MM - US
Common Device Name
CATHETER, INTRAVASCULAR OCCLUDING, TEMPORARY
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key14675022
Report Number3008881809-2022-00289
Device Sequence Number12115104
Product Code MJN
UDI-Device Identifier07613327004694
UDI-Public07613327004694
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K122576
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/13/2024
Device Model NumberM003SSC04070
Device Catalogue NumberM003SSC04070
Device Lot Number22644195
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/19/2022
Supplement Date Received by Manufacturer08/11/2022
Initial Report FDA Received Date06/13/2022
Supplement Report FDA Received Date09/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/14/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-